5618B Small Diameter Industrial RTD

For secondary level performance with full ITS-90 calibration, Fluke Calibration’s 5618B series PRTs are an excellent choice for temperature measurements

 

Secondary level performance with full ITS-90 calibration

  • Small diameter sheath, 3.2mm (0.125 in)
  • Excellent stability
  • Includes ITS-90 coefficients
  • NVLAP-accredited calibrated from –200°C to 500°C, lab code 200706-0

For secondary level performance with full ITS-90 calibration, Fluke Calibration’s 5618B series PRTs are an excellent choice for critical temperature measurements. Featuring a 3.2 mm diameter (1/8 in) sheath, these industrial standards probes have reduced response time without compromising precision. This small diameter 5618A probe works well in many applications where immersion depth is limited. Larger diameter probes give more measurement error in short immersion depth applications because they conduct more heat between ambient and the sensor.

With each probe you will receive a full NVLAP-accredited calibration report. On the report you’ll get the test data and the ITS-90 calibration coefficients that you can easily input into your Fluke Calibration thermometer.

The 5618B is also a great probe to use for calibrating your Fluke Calibration 9132 or 9133 infrared calibrators. In fact, these IR black body heat sources were designed to be calibrated with this type of probe. Now you can calibrate these targets in your own lab!

For use from –200 °C to 500 °C (the six-inch model goes to 300 °C), you won’t find a better industrial standard in this configuration than our 5618B. We recommend using the 5618B PRTs with the 1523, 1524, 1502A, 1529, or 1560 thermometer readouts.

 

 

 

Specifications

Resistance Nominal 100 Ω at 0 °C
Temperature
Coefficient
0.003923 Ω/Ω/ °C nominal
Temperature
Range
–200 °C to 500 °C
(–200 °C to 300 °C for 5618B-6-X)
Drift Rate ± 0.1 °C when used periodically to 500 °C
Sheath Material 316 SST
Leads 22 AWG PTFE, 6’
Termination Specify
Hysteresis Less than 0.01 °C at 0 °C when using –196 °C and
420 °C as the end points.
Time Constant Four seconds maximum for 63.2 % response to
step change in water moving
at 3 fps.
Thermal EMF Less than 25 mV at 420 °C
Calibration Includes manufacturer’s NVLAP-accredited
calibration w/ITS-90 coefficients, R vs. T values in
1 °C increments, lab code 200706-0
Size 5618B-12: 305 mm L x 3.2 mm dia. (12 x 1/8 in)
5618B-9: 229 mm L x 3.2 mm dia. (9 x 1/8 in)
5618B-6: 152 mm L x 3.2 mm dia. (6 x 1/8 in)
Calibrated
Accuracy† (k=2)
± 0.026 °C at –200 °C
± 0.066 °C at 0 °C
± 0.219 °C at 500 °C
†Includes
calibration
uncertainty and
drift specification. 
NVLAP† Calibration Uncertainty
Temperature Expanded
Uncertainty
(k=2)
–196 °C 0.024 °C
–38 °C 0.011 °C
0 °C 0.010 °C
200 °C 0.018 °C
420 °C‡ 0.029 °C
Note: Calibration uncertainties depend on the uncertainties of the lab performing the calibration. Subsequent calibrations of this probe performed with different processes, at different facilities, or with changed uncertainty statements may state different uncertainties. †Lab code 200706-0 ‡300 °C for 5618B-6
 

 

 

Model Name

Description

5618B-12-X

305 mm (12 in) Small Diameter Probe

(17025 accredited calibration included. Traceable to NIST standards.)

X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “I" (INFO-CON for 1521 or 1522), “J" (banana plugs), “L" (mini spade lugs), "M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs).

5618B-9-X

229 mm (9 in) Small Diameter Probe

(17025 accredited calibration included. Traceable to NIST standards.)

X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “I" (INFO-CON for 1521 or 1522), “J" (banana plugs), “L" (mini spade lugs), "M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs).

5618B-6-X

152 mm (6 in) Small Diameter Probe

(17025 accredited calibration included. Traceable to NIST standards.)

X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “I" (INFO-CON for 1521 or 1522), “J" (banana plugs), “L" (mini spade lugs), "M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs).

 

Accessories common to all models:

   

Accessory

Description

2601 Probe Carrying Case, Plastic (for probes up to 305 mm [12 in] in length)
   

 

Request a quote - Shop

  • Any additional requirements? Please let us know.
  • This field is for validation purposes and should be left unchanged.

Call Now Button