Smart Control Valve Testing is now easier than ever
The Fluke 710 Valve Testing Loop Calibrator is designed to enable users to perform quick, easy tests on HART smart control valves. Featuring built-in test procedures and an intuitive user interface, the 710 allows users to quickly and easily perform valve tests, while the valve test quick-check results provide at-a-glance diagnostics help you make maintenance decisions faster than ever. The valve health quick-check results let you know whether your valve is in good, marginal or in bad operating condition so you can quickly decide whether additional maintenance is necessary.
Valve testing and HART communication in a precision loop calibrator
With the 710 Valve Testing Loop Calibrator’s built in HART communication function, users can source a 4-20 mA signal to cause the smart control valve to move, while simultaneously interpreting the valve’s HART feedback signal to determine whether the valve is moving to the expected position. In addition to positional information, the measured pressure delivered from the valve’s internal I/P (which moves the valve) can be determined through the HART communication protocol.
The 710 has built-in test procedures that automatically increase and change the mA signal while monitoring the HART position and pressure feedback from the control valve, giving you a better overall picture of valve health at the simple push of a button.
Pre-configured valve tests, at-a-glance answers
Valve test routines built into the 710 include:
ValveTrack™ Software enables further analysis and trending
Valve tests that are logged and recorded to memory in the 710 are available to upload to included ValveTrack™ analysis software.
ValveTrack™ software enables you to:
Saving time, getting answers
In addition, the 710 offers:
Product Highlights
HART communication
The Fluke 710 offers a built-in HART modem to communicate the following HART commands:
HART Commands for Valves
The 710 includes these unique HART commands to support control valves:
In addition, the Fluke 710 offers:
Detailed Specifications |
|
Functions | mA source, mA simulate, mA read, mA read/loop power, and volts read. |
Ranges | mA (0 to 24 mA) and Volts (0 to 30 V DC) |
Resolution | 1 µA on mA ranges and 1 mV on voltage range |
Accuracy | 0.01% ± 2 counts, all ranges (@ 23 °C ± 5 °C [73.4 °F ± 41 °F]) |
Operating temperature range | -10°C to 55°C (14°F to 131°F) |
Humidity range | 10 to 95% non-condensing |
Stability | 20 ppm of F.S. /°C from -10°C to 18°C and 28°C to 55°C |
Display | 128 x 64 pixels, LCD Graphic w/backlight, .34 high digits |
Power | 6 AAA alkaline |
Battery life | 40 hours continuous use (measure mode using) |
Loop compliance voltage | 24 V DC @ 20mA |
Over-voltage protection | 240 V AC |
Overload current protection | 28 mA DC |
EMC | EN61326 Annex A (Portable Instruments) |
Dimensions (L x W x D) | 15 x 9 x 3 cm (6 x 3.6 x 1.3 in) |
Weight | 0.3 kg (9.5 oz) |
Included accessories | Traceable calibration certificate with data, batteries, test leads, USB upload cable and safety |
Warranty | Three year |
Includes: