Fluke 732B Direct Voltage Standard

The 732B is a direct voltage standard with 10V and 1.018V outputs.

 

The 732B is a direct voltage standard with 10V and 1.018V  outputs.

The 734A is a direct voltage reference standard that consists of four  732Bs that are mechanically and electrically isolated, housed in a  rack-mountable enclosure.

Each standard is small, light and rugged, ideal for shipment.

The long 72-hour battery life allows the 732B to be shipped over long distances under power. An optional external battery and charger extends battery life still further, to more than 130 hours.The 732B can stand up to a lot of abuse. The inputs can be shorted indefinitely and are protected up to 1100V dc, 25 mA, without damaging the cell or affecting its output.

Confidence

The 732B is based on the proven technology of the 732A, the first standards lab quality electronic reference to gain acceptance as a replacement for saturated standard cells. Originally designed for internal transfers of the Fluke Calibration corporate volt to the production floor, thousands are now in service world-wide in a variety of applications -- from maintaining an institutional reference to transferring values from national labs or privately-operated 10V Josephson Arrays.

Stability

Stability for each output is ±2 ppm per year (10V) and ±0.8 ppm per month (1.018V). Each 732B can drive up to 12 mA of current to simplify use with instrumentation - like the 5700A - with low output impedance.

Support for artifact calibration

Combined with the 742A-1 and 742A-10k Resistance Standards, a single 732B makes a tough and compact artifact calibration support package for instruments like the 5700A/5720A Calibrators from Fluke Calibration or the 3458A Multimeter from Hewlett-Packard.

Traceability

732Bs can be ordered with or without calibration certificates. The standard 732B comes without any formal traceability or any calibration certificate. Optionally a calibrated 732B can be ordered and requires both shipping in a “power on” condition (there are limits for the overall shipping transportation time). The model 732B/H is a 732B which ships with a Report of Calibration showing the deviation from nominal and the uncertainty at the time of the calibration. The 732B/C is similar to the 732B/H, but it adds a drift rate characterization report. This is a 12 month projection into the future from the time of calibration of the changes in the calibrated output value. Normally, the 732B calibration certificates are traceable to international standards but are not considered accredited calibrations. For an extra charge, an accredited calibration certificate is available. Contact your Fluke Calibration representative for details.

 

 

 

Output

10V
30 Day Specification:  ± 0.3 ppm
1 Year Specification:  ± 1.6 ppm
1.018V
30 Day Specification:  ± 0.8 ppm
1 Year Specification:  n/a

General Specifications

Temperature
 15°C to 35°C, operating; -51°C to 71°C, non-operating (with internal battery pack switched off)
Relative Humidity
 <95% to 30°C, <75% to 35°C, non-condensing
Altitude
 <3049m (10,000 ft) operating, <12,195m (40,000 ft) non-operating
Vibration
 Per MIL-T-28800; Type III, Class 5, Style E
Safety
 Designed to IEC 348, 2nd edition; 1978 and ANSI/ISA-S82; and UL1244, 2nd edition 1980, CSA C22.2 No 231, and IEC 1010
Calibration Documentation Reports of Calibration are optional
Size
 --------
732B DC Standard:  13.4 cm H x 9.8 cm W x 40.6 cm D (5.28 in H x 3.86 in W x 15.98 in D)
 --------
732B-7001 External Battery and Charger:  13.4 cm H x 9.8 cm W x 40.6 cm D (5.28 in H x 3.86 in W x 15.98 in D)
Weight
 --------
732B DC Standard:  5.91 kg (13 lbs.)
 ---------
732B-7001 External Battery and Charger:  5.45 kg (12 lbs.)
Power, battery life
732B DC Standard:  100V, 120V, 220V, 240V ac ± 10%; 50 Hz to 60 Hz; 66W max. Internal 12V lead acid, gelled electrolyte battery operates for 70 hours at 23°C when fully charged. Trickle-charged continually when external ac power is applied.
 --------
732B-7001 External Battery and Charger:  100V, 120V, 220V, 240V ac ± 10%; 50 Hz to 60 Hz; 66W max. Internal 12V lead acid, gelled electrolyte battery operates for 70 hours at 23°C when fully charged. Trickle-charged continually when external ac power is applied.

 

 

Model Name

Description

732B DC Standard, without calibration certificate, shipped under a power-off condition
732B/H Standard. With calibration certificate included, shipped under power
732B/C Standard with drift characterization and calibration certificate included, shipped under power

 

Request a quote - Shop

  • Any additional requirements? Please let us know.
  • This field is for validation purposes and should be left unchanged.