Fluke 5500A Multi-Product Calibrator

The 5500A is a revolutionary product that addresses a wide cross-section of your electrical calibrationwork load.

 

Technical Specifications

Voltage DC
Range:  0 to ±1020V
Best 1 Year Specification:  ±50 ppm of setting
Current DC
Range:  0 to ±11A
Best 1 Year Specification:  ±0.01% of setting
Resistance
Range:  0 to 3.29.999 MΩ
Best 1 Year Specification:  ±0.009% of setting
Voltage AC
Range:  1 mV to 1020V
   10 Hz to 500 kHz, Sine
Best 1 Year Specification:  ±0.03% of setting
Current AC
Range:  29 m A to 11A
   10 Hz to 10 kHz, Sine
Best 1 Year Specification:  ±0.06% of setting
Capacitance
Range:  0.33 nF to 1.1 mF
Best 1 Year Specification:  ±0.25% of setting
Thermocouple Source and Measure
Range:  -250°C to +2316°C (11 types)
Best 1 Year Specification:  ± 0.14°C of setting
RTD Source
Range:  -200°C to +630°C
   (8 types)
Best 1 Year Specification:  ±0.03°C of setting
DC Power
Range:  109 mW to 11 kW
Best 1 Year Specification:  ± 0.08% of setting
AC Power
Range:  109 mW to 11 kW
   45-65 Hz, PF=1
Best 1 Year Specification:  ± 0.15% of setting
Phase
Range:  0 to ± 179.99°
Best 1 Year Specification: ±0.15°
Frequency
Range:  0.01 Hz to 2.0 MHz
Best 1 Year Specification:  ±25 ppm of setting

General Specifications

Warm up Time
   Twice the time since last warmed up, to a maximum of 30 minutes
Settling Time
   Less than 5 seconds for all functions and ranges except as noted
Standard Interfaces
   IEEE-488 (GPIB), RS-232, 5725A (5500A only)
Temperature Performance
Operating:  0°C to 50°C
Calibration:  (tcal): 15°C to 35°C
Storage:  -20°C to 70°C
Temperature Coefficient
   Temperature coefficient for temperatures outside tcal ± 5°C is 10% of the 90-day specification (or 1-year, as applicable) per °C
Relative Humidity
Operating:  < 80% to 30°C, < 70% to 40°C, < 40% to 50°C
Storage:  < 95%, non-condensing
Altitude
Operating:  3,050m (10,000 ft) maximum
Non-operating:  12,200m (40,000 ft) maximum
Safety
   Designed to comply with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994; CAN/CSA-C22.2 No. 1010.1-92
Analog Low Isolation
   20V
EMC
   Designed to comply with FCC Rules Part 15
Line Power
Line Voltage (selectable):  100V, 120V, 220V, 240V
Line Frequency:  47 to 63 Hz
Line Voltage Variation:  : ± 10% about line voltage setting
Power Consumption
   300 VA
Calibration Documentation 17025 accredited report of calibration included
Dimensions
Height:  17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet on bottom of unit
Width:  43.2 cm (17 in), standard rack width
Depth:  47.3 cm (18.6 in) overall
Weight (without options)
   20 kg (44 lbs.)
Absolute Uncertainty Definition
   5500A and 5520A uncertainty specifications include stability, temperature coefficient, linearity, line and load regulation, and the traceability of the external standards used for calibration. You do not need to add anything to determine the total uncertainty of your calibrator for the temperature range indicated.
Specification Confidence Interval
   > 99%

 

 

Model Name

Description

5500A Multi-Product Calibrator

 

Accessories common to all models:

   

Accessory

Description

55XX/CASE Transit case with wheels
5500A/COIL 50-Turn Current Coil
5500A/HNDL Side handle
5500A/LEADS Test Lead Set
5500A-SC300 300 MHz Oscilloscope Calibration Option
5500A-SC600 600 MHz Oscilloscope Calibration Option
5725A Amplifier
Y5537 Rack mount kit
Priority Gold CarePlan - Electrical Comprehensive instrument calibration and repair support plans are available for this product

 

      Brochures

 

      Data sheets

 

 

Request a quote - Shop

  • Any additional requirements? Please let us know.
  • This field is for validation purposes and should be left unchanged.

Call Now Button