P5523 Liquid-to-Gas Media Separator

For high pressure pneumatic calibrations this unit interfaces with hydraulic deadweight testers providing a rateless liquid-to-gas separation.

Pressure range 0 to 3,000 psi (210 bar)

For high pressure pneumatic calibrations this unit interfaces with hydraulic deadweight testers providing a rateless liquid-to-gas separation. Driven by a pneumatic pressure source (nitrogen bottle) and controlled by needle valves, high pressure gas is balanced against the liquid in a hydraulic deadweight tester through a series of fluid traps.

* Note---since fluid mist may be transferred during operation we do not recommend this device for oxygen safe instruments using an oil operated deadweight tester.

 

 

 

Specifications

Instrument Size (W x D x H) 17.5 x 12 x 8.5 in ( 440 x 300 x 215 mm)
Instrument Weight 24 lb (11 kg)
Test Port Adapters 2 sets, 1/8, 1/4, 3/8 and 1/2 NPT or BSP
Gas Supply Connection 1/4 NPT female

 

 

Model Name

Description

P5523 Liquid to Gas Media Separator

Accessory Description
P5523 Liquid-to-Gas Media Separator Pressure range 0 to 3,000 psi (210 bar)

For high pressure pneumatic calibrations this unit interfaces with hydraulic deadweight testers providing a rateless liquid-to-gas separation. Driven by a pneumatic pressure source (nitrogen bottle) and controlled by needle valves, high pressure gas is balanced against the liquid in a hydraulic deadweight tester through a series of fluid traps.

* Note---since fluid mist may be transferred during operation we do not recommend this device for oxygen safe instruments using an oil operated deadweight tester.

 

      Product manuals

 

      Data sheets

 

Request a quote - Shop

  • Any additional requirements? Please let us know.
  • This field is for validation purposes and should be left unchanged.

Call Now Button