Tektronix 7904 Oscilloscope

The Tektronix 7904 is a 500 MHz, non-storage 7000-series oscilloscope mainframe with four bays. It accepts two 7000-series vertical plug-ins and two 7000-series horizontal plug-ins. There is also a rack-mount version, the R7903, albeit with only one horizontal bay.

The 7904 was introduced in late 1971/early 1972 (Note 1). In 1984, it was replaced by the redesigned 7904A.

Performance is largely determined by the plug-ins. The 7A19 vertical amplifier and 7B92 time base were introduced along with the 7904 to match the 500 MHz system bandwidth

Key Specifications

Bandwidth 500 MHz (with 7A19 or 7A29 amplifier)
Fastest calibrated sweep 500 ps/Div (with 7B92A, other time bases “uncalibrated for first 60 ns”)
Outputs Vertical Signal Out (25 mV/Div into 50 Ω, 0.5 V/Div into 1 MΩ); +Sawtooth; +Gate; camera power; probe power
Inputs Z Axis (15 Vp-p)
CRT
  • Standard: (8×10 Div @ 1 cm), P31 (green) 154-0644-05 or P11 (blue) 154-0644-09, acceleration voltage 24 kV (+21 / –3 kV)
  • Option 4, high writing speed: (8×10 Div @ 0.5 cm) , P31 (green) 154-0661-05 or P11 (blue) 154-0661-09
Calibrator DC, square wave (1 kHz or ½ B Sweep gate), 4 mV to 40 V in decade steps (2/20/200/400 mV into 50 Ω); 40 mA current loop
Power 90−132 V or 180−264 V, 48−440 Hz, 190 W max.
Weight 13.5 kg (30 lbs)
Dimensions 34.2 cm × 30.5 cm × 69.5 cm (13.5 in x 12.0 in x 23.8 in) H×W×L
Options
  • Opt.1: deletes read out system
  • Opt.2: X-Y delay compensation network
  • Opt.3: Extra EMI shielding
  • Opt.4: Maximum-brightness CRT with reduced area
  • Opt.78: P11 phosphor

 

Tek-7904-front