Tektronix FG502 calibration and repair

Home → Calibration Service → Electrical calibration → Tektronix → Tektronix FG502 calibration and repair

Have your Tektronix equipment calibrated in the  ISO/IEC 17025 accredited lab!

Minerva is specialized in repair and calibration services of Tektronix instruments. These Tektronix instruments are calibrated in a ISO/IEC 17025 accredited calibration lab.  

The calibration is performed by calibration technicians. We have more then 30 years of experience in calibrating instruments. In short, your Tektronix instrument is taken good care off! We focus on a fast lead time and extremely high standard of our service.

Tektronix FG502 Function Generator Plug-in

The Tektronix FG502 is a function generator plug-in for the TM500 system.

It generates low distortion sine, square, triangle, positive and negative pulse, and ramp waveforms from 0.1 Hz to 11 MHz.

A squarewave trigger is available at the front panel. Variable DC offset is also provided.

Α voltage-controlled frequency (VCF) input controls the output frequency from an external voltage source. The output frequency can be swept above or below the selected frequency to maximum of 1000:1 depending on the polarity and amplitude of the VCF input and the selected output frequency.

An external gate input allows the generator to operate for the duration of an externally applied gating signal. This mode provides either a single cycle output, or a train of preselected waveforms, depending on the gating signal width and the generator frequency setting.

Key Specifications

Frequency 0.1 Hz to 11 MHz (Sine, Square, Triangle), 0.01 Hz to 1.1 MHz (Ramp, Pulse), resolution 10-4
Frequency stability ≤0.1% for 1 hour, ≤0.5% for 24 hours (ambient +25±5°C)
Frequency accuracy ≤3% of full scale from 0.1 Hz to 1 MHz. ≤5% of full scale from 1 MHz to 10 MHz
Output Amplitude ≥10 Vp-p open circuit. ≥5 Vp-p into 50 Ω (@ 10 kHz)
Output Offset ≥±5 Vp-p open circuit. ≥±2.5 Vp-p into 50 Ω
Amplitude flatness Sine, ±1.5 dB ref. to 10 kHz; Square, triangle ±3 dB referenced to sine
Sine Distortion ≤0.5% from 10 Hz to 50 kHz. Harmonics ≥30 dB down at all other frequencies
Triangle Symmetry and Linearity Within 1% from 0.1 Hz to 1.1 MHz, 3% from 1.1 MHz to 11 MHz
Risetime ≤20 ns (square/pulse)
VCF 10 V signal shifts frequency ≥1000:1
Ext. Gate 0 to ≥+2 V, not to exceed +15 V
Trigger Output ≥5 V open circuit, ≥2.5 V into 50 Ω
Functions Sine, Square (50%), Triangle, + and − Ramp, + and − Pulse (5/95%)
tektronix_fg502